Olympia AN60-1

Nitrogen and Air Generator
Designed to supply LCMS and MP-AES analyzers with Nitrogen and Air

APPLICATIONS

– For single or multiple installations of LCMS, MP-AES, and all analyzers requiring oil-free dry air and nitrogen

Description

Delivers dry air and nitrogen up to 70 Nl/min (combined total flow) at pressures up to 9.5 bar
Low-noise operation thanks to advanced soundproofing and anti-vibration design
Equipped with an oil-free Scroll compressor
Maximizes energy efficiency through inverter technology, automatically adjusting power output according to demand
Single-phase power supply: 230 Vac, 60 Hz
Nitrogen and air pressure dew point: -20°C PDP (ATP)
Equipped with two 15-liter internal air receivers, ensuring stable flow even at full load
Microprocessor-based control system with integrated full-color touchscreen display
Compatible with MODBUS and MQTT communication protocols
Network connectivity via RJ45 Ethernet and/or Wi-Fi

Technical specifications

Specification

Value

Output Flow Rate

Dry Air + Nitrogen up to 70 Nl/min

Compressor Type

Oil-free Scroll compressor

Output Pressure

6–9.5 bar

Power Supply

230 Vac ±10%, 60 Hz, 2.5 kW

Protection Rating

IP20

Noise Level

Minimum: 54 dBA • Typical: 58 dBA

Operating Temperature

5 to 35°C

Storage Temperature

5 to 50°C

Ambient Humidity

Max. 90% RH, non-condensing

Maximum Altitude

Tested up to 2,000 m above sea level

Dimensions (W × D × H)

68 × 106 × 80 cm

Weight

155 kg (330 lbs)

Heat Dissipation

2340 J/s

Pneumatic Connections

Nitrogen outlet: G 1/2" female; condensate drain: Ø9 mm PVC tube with collection container

Electrical Connections

5 m power cable; IEC309 blue plug, 2 poles + earth, 32 A, IP44, compliant with IEC 60309-1 and IEC 60309-2

Data Connection

Ethernet communication via MQTT and MODBUS; RJ45 Ethernet or Wi-Fi connection

Warranty

2 years

 

 
 
 

 

 